On Thu, 29 Apr 2021 00:29:48 +0200, Sebastian Reichel wrote:
This series adds support for another General Electric patient
monitor series (similar to existing Bx50v3), which is based on
i.MX6DL using Congatec's QMX6 module.

The module uses an I2C RTC to provide the i.MX6 32768 Hz clock,
so it's important to keep it enabled. Not doing so results in
incorrect timings of watchdog and i.MX6 RTC. The bootloader
enables the watchdog, so disabling the clock results in system
reboot. [0]

[...]
Applied, thanks!

[1/5] rtc: m41t80: add support for fixed clock
      commit: f765e349c3e1f2e676ad4bd61197216b26976022

Best regards,
-- 
Alexandre Belloni [off-list ref]